Hardware design of sphygmomanometer system

文章來源:鑫志科技 / 作者:鑫志科技 / 發表時間:2021-09-14
Overall system structure

The system hardware mainly includes: LM3S1138 controller, air pump, filtering and amplification circuit, alarm circuit, storage module, keyboard module, and LCD display module.

The pneumatic circuit controls external pressure, collects pressure signal data, amplifies and filters the pulse signal in the mixed blood pressure signal, and amplifies the pulse signal. The storage module is mainly used to store measurement records, and the alarm circuit is used for displaying sound and light alarms in abnormal situations; Display module, used to display blood pressure measurement data and blood pressure history records; The keyboard is used to input system setting parameters.

Pneumatic circuit:

The pneumatic circuit consists of a cuff, an air pump, a pressure sensor, and an exhaust valve, all of which form a four connected structure. The two signals from the pressure sensor are connected to the two A/D channels of the main controller to measure the static pressure signal and pulse signal, inflation and deflation to complete the measurement.

Pressure sensor:

The physiological signals of the human body have the characteristics of low frequency and low signal, usually μ V~mv, with a frequency range of 0~300 Hz and a low signal-to-noise ratio, it is important to choose a sensor with good performance.

The MPXVS050GP pressure and internal control of the sensor enterprise includes a signal operational amplifier and signal regulation system function, which has well-developed linearity. The output working voltage is proportional to the applied pressure, and can effectively compensate on chip. Its temperature compensation characteristics can overcome the temperature drift problem in semiconductor pressure sensitive devices. It can directly convert the blood pressure on the arterial wall to 04.7V electrical signal, and the corresponding blood pressure value is 0375 mmHg, which is very consistent with the design requirements of sphygmomanometer, especially suitable for oscillographic pressure measurement.

Driver circuit design:

Air drive circuit. The driving currents of the Hlpm30a air pump and the hlv01a solenoid valve are 450ma and 75ma, respectively. The signals controlling the operation of the air pump and solenoid valve are sent by the lm3s1138 controller, but the digital I/O output current of the lm3s1138 controller cannot meet the requirements. Therefore, Uln2803 is provided for the air pump and solenoid valve to output a 500ma current for driving the air pump and solenoid valve. The first and second circuits of Uln2803 are used to drive electromagnetic valves and pumps, as well as filtering and amplification circuits, respectively.

The signal from the pressure sensor is a mixed signal of pulse signal and static pressure signal, with high-frequency interference and DC or low-frequency components from the outside. Static pressure signal is a low-frequency signal with a frequency less than or equal to 0.04 Hz, and the frequency of pulse signal is generally around 1 Hz. Although the selected pressure sensor has amplification power, the amplification technology for pulse oscillation frequency signals is limited, mainly due to static pressure signals. The mixed control signal can be divided into the following two parts and connected to the LM3S1138 controller. The development of one part directly affects the operation of the A/D converter for analog-to-digital conversion, and obtains static pressure signal analysis data; The other is to amplify the pulse signal data by analog-to-digital conversion after passing through the band-pass filter amplifier module circuit.

The filter consists of two RC networks that determine the cutoff frequency. The two cutoff frequencies are calculated by the following formula.

Considering the occurrence of hypotension and the fact that the patient's blood pressure and heart rate are lower than those of normal individuals, as well as the ability to filter out system interference signal components, we set the development pulse wave frequency to be 0.5-5 Hz. Therefore, R1=1 k Ω, C1=33 μ F,R3=1 MΩ,C2=0.33 μ F。

This circuit can not only extract the required pulse signal, but also amplify weak pulse signals. The strength of the pulse signal varies from person to person, but the general range is 1 to 3 millimeters of mercury. According to the transfer function mpxv5050gp of the pressure sensor, the voltage signal converted from this pressure signal is 12-36mv. Because of filtering

The oscillator attenuates by 10 dB at a 1 Hz signal, resulting in an oscillation wave signal of 3.8-11.4 mV. After testing, the amplification factor of the amplifier is set to 150, so that the signal variation range is located in the area with high A/D conversion sensitivity of LM3S1138, which can improve the accuracy of data acquisition.

咨詢

QQ

在線客服

服務時間:9:00-18:00

電話

服務咨詢熱線:

0755-2780-8770

微信

二維碼

微信咨詢

郵箱

電子郵箱

xuzb@xzkj.net.cn

主站蜘蛛池模板: 亚洲欧美色综合自拍 | 中文精品久久久久国产网站 | 日韩欧美中文字幕一区 | 99热这就是里面只有精品 | 久久99精品久久久久久国产越南 | 一级肉体毛片视频免费看看 | 九九精品视频免费 | 视频在线观看一区 | 美女一级免费毛片 | 日韩中文字幕在线观看视频 | 中文视频在线 | 日韩字幕无线乱码 | 精品国产免费第一区二区三区日韩 | 奇米影视777狠狠狠888不卡 | 亚洲永久视频 | 香蕉视频免费看 | 日韩亚洲综合精品国产 | 一区二区三区国产精品 | 日韩欧美亚洲每日更新网 | 日韩 欧美 亚洲 | 3www黄| 精品日产一区二区三区手机 | 国产成人爱片免费观看视频 | 91精品啪在线观看国产老湿机 | 在线视频一二三区2021不卡 | 女孕学护士一级毛片 | 色偷偷91久久综合噜噜噜 | 一级午夜视频 | 天天操天天操天天操香蕉 | 好吊顶色| 色姑娘天天操 | 狠狠综合欧美综合欧美色 | 怡红院免费va男人的天堂 | 好看的中文字幕在线 | 午夜欧美激情 | 国产第九页 | 亚洲欧美日韩精品久久久 | 国产在线精品二区赵丽颖 | 九色国产在视频线精品视频 | 9久热这里只有精品视频在线观看 | 手机在线观看亚洲国产精品 |