當前位置:首 頁 ? News ? Company News ? Design considerations of sphygmomanometer

Design considerations of sphygmomanometer

文章來源:鑫志科技 / 作者:鑫志科技 / 發表時間:2021-09-09

The blood pressure monitor or sphygmomanometer mainly uses an inflatable cuff and monitoring equipment to measure blood pressure through listening; Pressure sensors can also be used to measure blood pressure. This monitor can be completed through two ways: manually inflate the cuff, and use a stethoscope to monitor the sound from the artery wall (stethoscopy); Or insert a pressure sensor inside the sphygmomanometer to stop measuring the vibration of the artery wall (vibration measurement method).

Automatic electronic sphygmomanometer

There are two types of automatic detection sphygmomanometer: upper arm sphygmomanometer and wrist sphygmomanometer. The upper arm sphygmomanometer includes an air bag, which is tied to the upper arm during measurement. The inflatable cuff is connected to the monitor placed near the left side of the arm through a leather tube. The size of wrist sphygmomanometer is small, and the whole installation can be wrapped in tricks. Therefore, this kind of product has strict requirements for size. Some upper arm sphygmomanometer require manual inflation, but most upper arm and wrist sphygmomanometer on the market at present are fully automatic.

Measuring Technology

The automatic electronic sphygmomanometer first inflates the cuff to make it have enough pressure, so as to prevent blood from moving in part of the aorta. As the pressure gradually releases, when blood begins to flow through the arteries, the measured pressure value is the systolic blood pressure of the heart. At the same time, it can also monitor the rate of pulse beating. When the blood flow returns to normal, the measured value is equivalent to the diastolic blood pressure of the heart. This completes a complete measurement cycle, which is completed by the installed pump, inflatable cuff, electronic valve, and pressure sensor.

The signal generated by the pressure sensor is processed by an operational amplifier or instrument amplifier and sent to an analog-to-digital converter (ADC). Then, for different types of monitors and sensors, appropriate methods are used to calculate systolic blood pressure, diastolic blood pressure, and pulse rate in the digital domain, and the resulting results are displayed on a liquid crystal display (LCD). Typically, the data and corresponding time/date are also stored in non volatile memory.

data interface

Some sphygmomanometer can upload data to the computer so as to stop further analysis and later measurement tracking, and can transmit data through USB interface. You can choose a discrete USB transceiver or a microcontroller with a built-in transceiver.

Audio prompts

The audio prompt of the sphygmomanometer can be a simple beep or a high-end voice prompt. A simple buzzer can be driven by one or two pulse width modulation (PWM) output port pins of a microcontroller; Regarding the advanced voice prompt function circuit, it is possible to select an audio digital to analog converter (DAC) and speaker amplifier to process audio signals.

Power management

The upper arm sphygmomanometer usually uses 4 AAA (1.5V) alkaline batteries for power supply, while the wrist sphygmomanometer mostly uses 2 AAA alkaline batteries for power supply. The pump and analog circuit of the sphygmomanometer request 5V or 3.3V power supply, and the digital circuit selects 3.3V or 1.8V power supply according to the measurement technology adopted in detail. Therefore, a typical upper arm sphygmomanometer needs a buck boost switching power supply to provide 5V power for the pump/analog circuit, and a low dropout linear regulator (LDO) to provide 3.3V power for the digital circuit. A typical wrist sphygmomanometer uses boost switching power supply to provide 3.3V power required by the pump/analog circuit, and uses an LDO to convert the input voltage to 1.8V to supply power for the digital circuit.

In order to extend the service life of the battery, when the sphygmomanometer is not working, the power supply can be turned off, and only the real-time clock (RTC) can be maintained effectively and the working state of the sphygmomanometer can be quickly restored.

Presentation and backlighting

Most sphygmomanometer use simple segment LCD displays, which contain 100 segments or less, and LCD drivers are usually integrated in the microcontroller. Apply one or two white light LED (WLED) or electroluminescent (EL) light sources to provide display backlight. The wrist sphygmomanometer can use a switching power topology to provide WLED drive, and the upper arm sphygmomanometer can use a linear regulator to provide WLED drive.

electrostatic discharge

All sphygmomanometer monitors must pass the electrostatic discharge (ESD) test request of IEC 61000-4-2, be able to apply devices with built-in ESD maintenance, or use ESD maintenance lines to maintain the lines exposed at the interface.

咨詢

QQ

在線客服

服務時間:9:00-18:00

電話

服務咨詢熱線:

0755-2780-8770

微信

二維碼

微信咨詢

郵箱

電子郵箱

xuzb@xzkj.net.cn

主站蜘蛛池模板: 久久精品中文 | 国产精品亚洲欧美一区麻豆 | 久久久久在线视频 | 欧美日韩国产高清 | 成人www | 狠狠色丁香婷婷综合视频 | 5g天天爽影院 | 色综合天天综合网看在线影院 | 99re热在线视频 | 中文字幕欧美在线观看 | 国产精品亚洲国产三区 | 五月天婷五月天综合网在线 | 男人的天堂久久精品激情 | 久久做 | 中文字幕在线观看不卡 | 久久婷婷人人澡人人爱91 | 国精品一区二区三区 | 欧美视频在线视频 | 97伦理片| 久久国产精品久久精 | 精品国产福利 | 亚洲精品高清国产一线久久97 | 丁香久久| 毛片天堂 | 夜色成人免费观看 | 亚洲精品不卡午夜精品 | 成人欧美在线 | 人人爱人人做 | 四虎影视久久 | 久草精品视频在线观看 | 国产 日韩 一区 | jizzjizz中国护士第一次 | 99热精品成人免费观看 | 国内一区亚洲综合图区欧美 | 99久久99久久精品免费看子 | 12345国产精品高清在线 | 国产小视频91 | 伊人五月天婷婷琪琪综合 | 亚洲免费一级视频 | 欧美成人毛片免费网站 | 四虎在线永久免费视频网站 |